Canine Anaplasma Antibody Rapid Test Cassette, Dog Test Kits, Veterinary Rapid Test Kits
Product: | Canine Anaplasma Antibody Rapid Test Cassette, Dog Test Kits, Veterinary Rapid Test Kits |
Specimen: | WB/S/P |
Shelf Life: | 2 Years |
Reading Time: | 10 Minutes |
Specificity: | 98.77% |
Principle: | Chromatographic Immunoassay |
Pack: | 10 T |
Format: | Cassette |
Sensitivity: | 95.45% |
Accuracy: | 98.06% |
INTENDED USE
The Canine Anaplasma Antibody (ANA Ab) Rapid Test Cassette (Whole Blood/Serum/Plasma) is a sandwich lateral flow immunochromatographic assay for the qualitative detection of Canine Anaplasma antibody (ANA Ab) in canine’s whole blood,serum or plasma.
PRINCIPLE
The Canine Anaplasma Antibody Rapid Test Cassette is based on lateral flow immunochromatographic assay. The test cassette has a testing window. The testing window has an invisible T (test) zone and C (control) zone. When sample is applied into the sample well on the cassette, the liquid will laterally flow on the surface of the test strip. If there are enough Canine Anaplasma antibodies in the sample, a visible T line will appear. The C line should always appear after a sample is applied, indicating a valid result. By this means, the cassette can accurately indicate the presence of Canine Anaplasma antibodies in the sample.
STORAGE AND STABILITY
Store as packaged in the sealed pouch either at room temperature or refrigerated (2-30 °C). The test is stable through the expiration date printed on the sealed pouch. The test cassette must remain in the sealed pouch until use. DO NOT FREEZE. Do not use beyond the expiration date.
DIRECTIONS FOR USE
Allow the test cassette, specimen, buffer to equilibrate to room temperature (15-30 °C) prior to testing.
1. Collect canine’s fresh whole blood, serum or plasma samples and ensure that the samples cannot be turbid or precipitated. Only use clear, non-hemolyzed specimens. Do not leave the specimens at room temperature for prolonged periods. Serum and plasma specimens may be stored at 2-8 °C for up to 3 days. For long term storage, serum and plasma specimens should be kept below -20°C.
2. Place the test cassette on a clean and level surface.
3. Hold the dropper vertically and transfer 2 drops of sample (approximately 20 μL) and immediately add 2 drops of buffer (approximately 80 μL) into the specimen well (S), then start the timer.
4. Read the result in 10 minutes. Do not interpret results after 15 minutes.